Specification
Product number
OSCSJ-Ⅱ-80L
Water inlet requirements
Municipal tap water, when working,
the inlet water pressure is 0.1-0.3Mpa, the water temperature is 15-35℃, and
the water quality meets the requirements of GB57492
System flow
PF+AC+RO+DI
Purification system
Precision filter*1+300GPD RO
reverse osmosis membrane*2+ion purification column*4
Resistivity
10-18.25MΩ.cm
heavy metal ion
<0.1ppb
bacterial
<0.1cfu/ml
particulates
<1/ml
Ion rejection rate
96-99%
Organic matter retention rate
>99%
Particle and bacteria retention rate
>99%
Water production (25℃)
80 liters/hour
Instant water output
2500ml/min
Outlet
1 RO reverse osmosis water
(optional), 1 DI deionized water (standard)
Use environment
Temperature: 15-35℃ Relative
humidity: ≤80%
Dimensions/weight
390*400*1000mm/35-45kg
Power supply/power
Rated voltage and frequency:
220V±10%, 50/60Hz, rated power: 230W
The standard configuration
Host + bucket + accessories