Specification
Product
number
OCSJ-Ⅱ-100L
Water inlet requirements
Municipal
tap water, when working, the inlet water pressure is 0.1-0.3Mpa, the water
temperature is 15-35℃, and the water quality meets the requirements of
GB57492
System flow
PF+AC+RO+DI
Purification system
Precision
filter*1+300GPD RO reverse osmosis membrane*2+ion purification column*4
Conductivity
0.055-0.1us/cm
Resistivity
10-18.25MΩ.cm
heavy metal ion
<0.1ppb
bacterial
<0.1cfu/ml
particulates
<1/ml
Ion rejection rate
96-99%
Organic matter retention rate
>99%
Particle and bacteria retention rate
>99%
Water production (25℃)
100
liters/hour
Instant water output
2500ml/min
Outlet
1
RO reverse osmosis water (optional), 1 DI deionized water (standard)
Use environment
Temperature:
15-35℃ Relative humidity: ≤80%
Dimensions/weight
390*400*1000mm/35-45kg
Power supply/power
Rated
voltage and frequency: 220V±10%, 50/60Hz, rated power: 230W
The standard configuration
Host
+ bucket + accessories