OLABO YTQ-F Electric Pneumatic Hemostat

Product Introduction

Automatic pneumatic tourniquet is a commonly used hemostatic instrument in the operating room. It is widely used in orthopedic surgery. Reasonable use can shorten the operation time, reduce the bleeding of the patient's operation area, make the operation field exposed clearly, and facilitate operation.


Features

The F-type hemostatic instrument has four alarm times, which can be set according to customer needs. The sequence of the four alarm times is arranged in the "upper left corner-upper right corner-lower left corner-lower right corner" window. Press the "alarm shape" key to set the alarm time in four windows in sequence. If there is no debugging for 5 seconds, the setting alarm time interface will jump to the working pressure and time debugging interface. Pay attention to the setting: the previous alarm time must be greater than or equal to the next alarm time, and the four alarm time must be less than the set minimum working time. The first four alarms are only three sound alarms, and the sound and light alarm will be sounded for a long time after the working time is over. The size of the alarm sound can be adjusted in 8 cycles by pressing the "horn shape" key. After manual deflation, the sound and light alarm ends, and the air pump of the instrument stops working, but the deflation solenoid valve is in a power-off state at this time, and the system pressure is in a natural holding state (there will be a slower natural pressure relief depending on the instrument conditions) .


Introduction of IVRA function of F type hemostatic instrument

After pressing the IVRA button, the pressure and working time of the right channel are controlled by the left channel. At this time, if the right channel starts to inflate before the set time of the left channel is over, the left channel window will have "DEF" flashing and Sound and light alarm prompts deflation. If the left channel starts to inflate when the set time of the right channel is not over, the right channel window will flash "DEF" and an audible and visual alarm to prompt deflation.
When the "IVRA" key is not pressed, the left and right channels can work separately without interfering with each other.
The F-type hemostatic instrument can provide internal power supply for the host to work. When the external power supply is not connected, you can press the "battery shape" key to query the power of the internal power supply. The power is displayed in the lower left corner of the window as a percentage. When the external power supply is connected, the internal power supply is automatically charged.

Technological Parameter

Rated working voltage 220VAC±22VAC
Input power Not more than 50VA
Maximum rated output gas pressure Not less than 90Kpa (675mmHg)
Inflation time for the tourniquet to reach the maximum rated gas pressure (from the pressure less than 50mmHg) ≤60s
Pressure adjustment accuracy ±10mmHg
Timing adjustment range 0-99 min (0-240min for F type)
Pressure display unit mmHg